The grading system shown in Table 11.2 is used throughout these guidelines for manual tube/microplate serological testing.
Table 11.2 Grading system for serological tests
Reaction grade | Description |
---|---|
Grade 5 |
Cell button remains in one clump or dislodges into a few large clumps |
Grade 4 |
Cell button dislodges into numerous large clumps |
Grade 3 |
Cell button dislodges into many small clumps |
Grade 2 |
Cell button dislodges into finely granular but definite, small clumps |
Grade 1 |
Cell button dislodges into fine granules |
Grade 0 |
Negative result |
Unless otherwise stated, an unequivocal manual tube reaction is defined as a grade 3 or greater and for column tests as 1+ or greater |
Performance evaluation should be undertaken in accordance with:
Stability testing should be performed in accordance with BS EN ISO 23640 In vitro diagnostic medical devices. Evaluation of stability of in vitro diagnostic reagents.
Where appropriate, the following requirements should also be included in performance evaluation:
Specificity tests
Requirements
Potency tests – tube or microplate methods
Table 11.3 Requirements for conventional blood typing reagents
Antibody specificity | Specification |
Performance evaluation (as a minimum, two examples of the following reference cells should be included if available)* |
Batch release testing | |||||
---|---|---|---|---|---|---|---|---|
Specificity | Potency | |||||||
Positive reactors | Negative reactors | |||||||
Cell type | No. | Cell type | No. | Cell type | No. | |||
anti-A |
Normally blue coloured Should equal or exceed potency of reference preparation(s) Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
AX,A3 |
A1 |
2 |
B |
2 |
See insert of reference preparation(s) |
|
A cord cells |
A2B |
2 |
O |
2 |
|
|
||
|
AX* |
|
|
|
|
|
||
anti-B |
Normally yellow coloured Should equal or exceed potency of reference preparation(s) Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
BX, B3, Bv |
B |
2 |
A1 |
2 |
See insert of reference preparation(s) |
|
B cord cells |
A1B |
2 |
O |
2 |
|
|
||
anti-A, B |
Normally clear coloured Should equal or exceed potency of reference preparation(s) Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
A1, A2, B, A1B, A2B |
A1 |
1 |
O |
4 |
See insert of reference preparation(s) |
|
AX A3 |
A2 |
2 |
|
|
|
|
||
Bx B3 |
B |
2 |
|
|
|
|
||
A and B cord cells |
AX |
2 |
|
|
|
|
||
anti-A1 |
Normally clear coloured Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
A1 |
2 |
A2 |
2 |
A1 |
2 |
|
A1B |
2 |
A2B |
2 |
|
|
||
|
|
|
B |
2 |
|
|
||
|
|
|
O |
2 |
|
|
||
anti-D |
Normally clear coloured Should equal or exceed potency of reference preparation(s) Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
Weak D (500 sites/cell) |
R1r |
2 |
r’r |
1 |
See insert of reference preparation(s) |
|
|
R2r |
2 |
r’’r |
1 |
|
|
||
DVI type 1, DVI type 3, DIV, DV, DVII, DFR, DBT, R0Har |
Weak D* |
2 |
rr |
1 |
|
|
||
anti-C |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
Cw, Cx, r’S |
R1r |
1 |
R2 R2 |
1 |
R1r |
2 |
R2RZ |
R1R2 R1RZ |
1 |
r’’r |
1 |
|
|
||
|
r’r |
1 |
rr |
1 |
|
|
||
anti-E |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
R1RZ |
R2r |
1 |
R1R1 |
1 |
R2r |
2 |
EW |
R1R2 |
2 |
r’r |
1 |
|
|
||
|
r’’r |
1 |
rr |
1 |
|
|
||
anti-c |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
R1RZ, R1wR1 |
R1r |
2 |
R1R1 |
3 |
R1r |
2 |
|
R1R2 |
1 |
|
|
|
|
||
|
r’r |
1 |
|
|
|
|
||
anti-e |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
R2RZ |
R2r |
2 |
R2R2 |
3 |
R2r |
2 |
|
R1R2 |
1 |
|
|
|
|
||
|
r’’r |
1 |
|
|
|
|
||
anti-Cw |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use |
R1wR1,r’wr, R1wr |
R1wr |
2 |
R1r |
1 |
R1wr |
2 |
|
R1wR2 |
2 |
R1R1 |
1 |
|
|
||
|
r’wr |
1 |
r’r |
1 |
|
|
||
anti-K |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
K+k+ Kp (a+b+) |
K+k+ |
4 |
K–k+ |
4 |
K+k+ |
2 |
K+k+ Kp (a–b+) |
|
|
|
|
|
|
||
anti-k |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
K+k+ Kp(a+) |
K+k+ |
4 |
K+k– |
4 |
K+k+ |
2 |
|
Kp(a+b+) |
2 |
|
|
|
|
||
|
K+k+ Kp(a–) |
2 |
|
|
|
|
||
anti-Fya |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
Fy(a+b+) |
4 |
Fy(a–) |
4 |
Fy(a+b+) |
2 |
anti-Fyb |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
Fyx |
Fy(a+b+) |
4 |
Fy(b–) |
4 |
Fy(a+b+) |
2 |
anti-Jka |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
Jk(a+b+) |
4 |
Jk(a–) |
4 |
Jk(a+b+) |
2 |
anti-Jkb |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
Jk(a+b+) |
4 |
Jk(b–) |
4 |
Jk(a+b+) |
2 |
anti-S |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
S+s–, |
S+s+ |
4 |
S–s+ |
4 |
S+s+ |
2 |
anti-s |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
S+s+ |
4 |
S+s– |
4 |
S+s+ |
2 |
anti-M |
Normally clear coloured Potency titre greater than 2 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
M–N+ He+ |
M+N+ |
4 |
M–N+ |
4 |
M+N+ |
2 |
anti-N |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
M+N+ |
4 |
M+N– |
4 |
M+N+ |
2 |
anti-P1 |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
P1+ strong |
4 |
P1– |
4 |
P1+ |
2 |
|
P1+ weak |
4 |
|
|
|
|
||
anti-Lea |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
Le(a+b–) |
4 |
Le(a–) |
4 |
Le(a+) |
2 |
anti-Leb |
Normally clear coloured Potency titre greater than 4 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
A1B Le(a–b+) |
A1B Le(a–b+) |
4 |
Le(b–) |
4 |
Le(b+) |
2 |
'Other' |
Normally clear coloured Potency titre greater than 2 vs by techniques detailed in the manufacturer’s instructions for use Should detect variants and subgroups as detailed in the manufacturer’s instructions for use |
|
Heterozygous positive |
4 |
Antigen negative |
4 |
Heterozygous positive |
2 |
* For reagents where reactivity against the antigen is claimed |
For batch acceptance testing the user must ensure that the typing reagent reacts with the weakest available antigen expressing cells (refer to batch release positive reactors in table 11.3) and does not produce false positives with cells negative for the antigen.
Monoclonal antibodies have been developed which necessitate revision of the optimal composition of anti-human globulin reagents. For example, because of the limitations imposed by the presence of C3d on normal red cells, particularly in stored blood, conventional polyclonal anti-complement reagents rely on anti-C3c to detect in vitro bound complement and limited amounts of anti-C3d to detect in vivo bound complement. However, some monoclonal IgM anti-C3d reagents can be used at concentrations adequate to detect both in vitro and in vivo bound complement without causing unwanted positive reactions with normal red cells and fresh, inert, group-compatible serum in routine tests.
Performance evaluation should be undertaken in accordance with:
Stability testing should be performed in accordance with:
Specificity testing
Tests for IgM or IgG red cell heterospecific antibodies
Requirements
Tests for unwanted positive reactions
Requirements
anti-IgG potency: polyspecific anti-human globulin and anti-IgG reagents for use in tube or microplate techniques
Requirements
Potency tests
anti-IgG potency by chequerboard titration studies with red cells sensitised with weak IgG antibodies (anti-D, anti-K and anti-Fya)
Details of tests are outlined in section 11.4.
Requirements
anti-complement potency; polyspecific anti-human globulin reagents for use in tube tests
Requirements
Tests for unwanted positive reactions
Instructions for use
The instructions for use for anti-human globulin reagents used in tube and microplate tests should also include a statement that:
Reagent red cells prepared from human blood are essential in ensuring safe transfusion practice. They are used in the determination of ABO blood groups, in the control of blood grouping reagents and of the anti-human globulin technique, and in the detection and identification of atypical red cell alloantibodies.
The immediate container and instructions for use sheet for reagent red cells should also meet the following criteria:
The detection of atypical antibodies in the serum of a patient is of greater clinical significance than if such antibodies are detected in blood donors. Reagent red cells of a lesser specification may be used when performing antibody screening tests on blood donor samples.
In general the following should apply:
C, c, D, E, e, K, k, Fya, Fyb, Jka, Jkb, S, s, M, N, P1, Lea and Leb
Fya, Fyb, Jka, Jkb, S and s
c, e, K, k, Fya, Fyb, Jka, Jkb, S, s, M, N, P1, Lea and Leb
Fya, Fyb, Jka, Jkb, S and s
D, C, c, E, e and K.
C, Cw, c, D, E, e, K, k, Kpa, Fya, Fyb, Jka, Jkb, S, s, Lea, Leb, M, N, P1 and Lua.
K, k, Fya, Fyb, Jka, Jkb, S and s.
c, k, Fya, Fyb, Jka, Jkb, S and s.
Reagent red cells for use in the identification of atypical antibodies should be confirmed as group O by an ABO blood grouping procedure which is capable of demonstrating the AX phenotype.
Reagent red cells
Reagent red cells for use in quantification techniques should be confirmed as group O by an ABO blood grouping procedure which is capable of demonstrating the AX phenotype.
Reagent red cells
These reagent red cells should be manufactured in accordance with the general guidelines in section 11.2.3.2.
When used in the formulation of reagents, fetal calf serum and bovine serum albumin should be obtained from a closed herd in the female line since 1980, in which no animal has been clinically suspected of having bovine spongiform encephalopathy (BSE), and which has not been fed rations containing ruminant-derived protein during that period.
Bovine albumin, usually supplied as a 20% or 30% solution, can be used as a constituent of a diluent for use in automated blood grouping antibody detection machines, for antibody quantification or as a potentiator in antisera, monoclonal reagents and anti-human globulin. When diluted and used in the system prescribed it should not cause:
The activity of each batch of proteolytic enzyme should be assessed to ensure batch-to-batch consistency using a biochemical assay (e.g. azo-albumin technique).5
For manual antibody detection techniques, red blood cells treated with the enzyme should achieve activity comparable to that of the reference enzyme preparation 92/658 used with an anti-D of 2.5 to 3.5 IU/mL.
For automated antibody detection techniques for patient pre-transfusion samples red blood cells treated with the enzyme should readily detect a weak anti-D of no more than 0.1 IU/mL (e.g. NIBSC anti-D standard for assessing operator and test performance as described at www.nibsc.org).
For automated antibody detection techniques for donation testing the red blood cells treated with the enzyme should readily detect a weak anti-D of 0.5 IU/mL.
The quality of water used in the production of a reagent should be adequate for that reagent. Ionic and non-ionic contaminants of water may interfere with components of reagents or may result in a conductivity or osmolality other than that intended. Water should have a conductivity of 1.0 µS/cm or less or a resistivity of 1.0 Mohm/cm or greater.
Saline is an isotonic solution containing 8.5 to 9.0 g/L NaCl (0.145–0.154 M) and should contain sufficient buffer to maintain pH 7.0 ±0.2 at 22 ±1°C during its shelf life.
The term low ionic strength solution (LISS) should not be used to denote a low ionic strength formulation other than that described by Moore and Mollison.6 LISS should not be used in place of preparations designed for a particular technology. LISS has the following properties:
The reactions obtained by an indirect antiglobulin test (IAT) with a weak anti-D and D positive cells suspended in LISS should be equal to, or better than, those obtained with the same cells suspended in saline and incubated at 37°C for 15 minutes.
Weak antibodies, such as anti-D, -K, -Fya can be used to control antibody detection techniques using indirect antiglobulin methods.
To act as a wash control the weak anti-D positive control could be diluted in serum or plasma. If the diluent is saline/bovine serum albumin, the control test could be positive, even though the cell washing was sub-optimal and this should be noted in the package insert.
These weak antibodies should:
For weak anti-D the antibody activity should be expressed in IU/mL.
As a minimum two vials containing red cells and plasma combined:
The reagent should be:
DTT can be used to alter the red cell membrane and/ or reduce the disulphide bonds of IgM molecules and can be supplied at different concentrations to treat red cells and plasma samples or reagents.
Fluorescently labelled monoclonal antibodies used as a group of reagents to accurately determine the size of an RhD positive fetal bleed in an RhD negative person.